Steam Autoclave 700-2000 Litre Capacity
Specially Designed for Hospital, Pharmaceutical and Laboratory Applications
• Installation with internal electrical steam generator or using central steam plant or double purposed
• Pneumatically operated single or double horizontal sliding door
• Complete sterilization cycles: pulse-matic, pre-vacuum, sterilization & post–vacuum
• With trolley for safe handling of wastes
• Touch screen LCD display for monitoring sterilizer’s operation
• PLC based control panel and all electrical components supplied from a certified international company
• Micro-biological filter for air refinement
• Pressure gauges for chamber, steam in-take & compressed air
• Steam traps for steam condensate outlet
• With water seal ring vacuum pump
• Chamber material: rustproof stainless steel; AISI SST 316 L
• Steam jacket material: rustproof stainless steel; AISI SST 304
• Structure material: rustproof stainless steel; AISI SST 304
• Front panel material: rustproof stainless steel; AISI SST 304
Steam Autoclave 700-2000 Litre Capacity
Specially Designed for Hospital, Pharmaceutical and Laboratory Applications
Supplied in three main models:
• Autoclaves using only central steam plant; appropriate for CSRs equipped with central steam plant.
• Autoclaves with internal electrical steam generator; appropriate for CSRs without central steam facilities.
• Double purposed autoclaves with a steam generator and capability of using central steam plant as well.
• Available in single or double door configuration.
• Double door autoclaves are supplied for central sterilization rooms with a separated sterile area which provides the most stringent clean room and safety conditions.
• The doors are automatic horizontal sliding doors, operated by a pneumatic mechanism.
• All doors are designed with a number of safety features.
• No door will open if there is still pressure in the chamber.
• No door will open if the chamber or load temperature is too high.
• In a double door system it is impossible to open two doors simultaneously.
• While door is being closed, the door will open immediately if an object is detected in front of it.